Energy Dispersive X-Ray Fluorescence (ED-XRF)Quant'X ARL, ThermoFisher
ED-XRFThermo Fisher ScientificARL Quant'X
Energy Dispersive X-Ray Fluorescence Spectrometry for non-destructive elemental analysis of solid materials, powders and finished objects
main features
- Silver (Ag) Target X-ray tube and High-Voltage power supply, Be window,50 W maximum output power (kV x mA), Voltage range 4 - 50 kV adjustable in 1 kV steps
- Silicon Drift Detector (SDD)
- 7 filters system, to acquire the signal from F to U
- Daily Energy Adjustment on Cu Standard
- Vacuum Analysis Option by Vacuum Pump (Edwards)
- Collimator Set 8.8mm, 3.6mm, 2mm, 1mm
- Camera for Sample Imaging XR-ED-CCD
- Gemstone pre-calibrated method (BX190409-2391): Corundums, Beryls, Chrysoberyl, Garnets, Spinels, Tourmalines and Pearls
- UniQuantTM Advanced Standardless FP for EDXRF
- Complete set of ThermoFisher Standards samples
General description
The laboratory is equipped with a Thermo Scientific ARL Quant'X spectrometer, a high-performance benchtop ED-XRF system for non-destructive elemental analysis.
The installed configuration allows qualitative and quantitative analysis over a wide range of elements, from Fluorine (F) to Uranium (U), with high sensitivity and operational flexibility.
Main technical features
- X-ray tube with Silver (Ag) target
- Maximum power: 50 W (kV × mA)
- Voltage range: 4 - 50 kV, adjustable in 1 kV steps
- Beryllium window (Be)
- High Resolution Silicon Drift Detector (SDD)
- 7 primary filter system for selective beam optimisation
- F to U analysis
- Collimator system: 8.8 mm, 3.6 mm, 2 mm, 1 mm
Optical system, detection and resolution
The SDD detector allows:
- High energy resolution
- Improved separation of elementary peaks
- Excellent sensitivity for trace elements
The system of 7 primary filters makes it possible to optimise excitation according to energy range and target element, improving signal-to-noise ratio and detection limits.
The system also provides for:
- Daily energy regulation on copper (Cu) standard to ensure stability and accuracy over time.
Configuration of accessories and measuring modes
The installed configuration includes:
- Option of vacuum analysis with Edwards pump, to improve sensitivity on light elements
- Imaging system with XR-ED-CCD chamber for visualisation and sample centring
- Complete set of interchangeable collimators to adapt the scanned area
- Complete set of ThermoFisher certified standards
- UniQuant™ Advanced Standardless FP software for quantitative analysis without specific standards.
Advantages of the technique
- Non-invasive and relatively quick technique;
- It allows qualitative, semi-quantitative and quantitative surveys)
- Elemental analysis from F to U
- It is possible to analyse samples from a few millimetres to whole objects up to a few tens of centimetres in size, as long as they are compatible with the size of the chamber;
- In several applications it is a key-test.